MODEL 2010/M PRISM COUPLER APPLICATION NOTES

BULK MATERIAL OR THICK FILM INDEX/BIREFRINGENCE MEASUREMENT


The Model 2010/M can operate as a fully-automated refractometer, providing high accuracy measurement of refractive index and index anisotropy for solid or liquid bulk materials and thick films in the index range 1.0 -2.6 without use of toxic or corrosive index matching fluids. This bulk material index measurement is not an option, but a standard feature provided with every system.

For bulk material index measurements, if a material with index n (see Fig. 1) is brought into intimate contact with a prism with index n
p.

Laser light directed onto the base of the prism will be totally reflected at the prism base until the angle of incidencebecomes less than the critical anglewhere

= arcsin(n/np) (1)

is easily measured using the detector in Fig. 1 since the detector intensity drops abruptly asdrops below the critical angle and light starts to leak into the bulk material. Since np is known, n can then be determined from Equation (1).

This same approach can also be used to measure refractive index for thick films. In the case of a film, the critical angle of the prism/film interface establishes an upper limit to the angle at which film propagation modes can occur. As a film with fixed index gets thicker, the angle at which the first propagation mode occurs asymptotically approaches the critical angle as defined by equation (1) where n now refers to the film index rather than to a bulk index. If we make the approximation that the angle of the first film mode equals the critical angle, the error in the measured index due to this approximation is less than .004 for a film thickness of 3 microns, less than .001 at 5 microns, and less than .0003 at 10 microns. Thus, this approach can be applied with good accuracy to measure the index of materials with thickness ranging from several microns on up to "bulk" thickness.

For measurement of liquid samples, a cell to bring the liquid in direct contact with the prism is available.

The bulk index approach can also be used to study index anisotropy in solid materials. In normal operation (Fig. 2a), the Model 2010/M uses a polarized laser with TE incidence, i.e., electric field vibrating transverse to the plane of incidence (out of the plane of Fig. 1). Thus, if the material in figure 1 is simply manually rotated against the prism (Fig. 2b), the electric field vector can be made to assume any orientation within the plane defined by the surface of the material, and the index for electric field vibration along any axis lying parallel to the material surface can be measured.



In addition, with option #2010/M-TM, which allows TM incidence (magnetic field transverse to, and electric field parallel to, the plane of incidence), index for electric field vibration perpendicular to the material surface is measurable (Fig. 2c). This is because the electric field vibrates perpendicular to the material surface at the critical angle since the refracted ray must propagate parallel to the surface at the critical angle.

Operation: Bulk index measurement is selected by depressing a function key (F9) on the control PC of the Model 2010/M. In the bulk index mode, the angle of incidence (È) is varied using a motor-driven rotary table until a sharp drop is detected on the detector. The system software then calculates and displays bulk index from equation (1).

Accuracy and resolution: With the high resolution rotary table (a no-cost option), worst case index accuracy is ±.001 and resolution is ±.0001 for bulk materials and films thicker than 10 microns. Absolute accuracy can be improved to approximately ±.0001, however, if the user is willing to perform a simple calibration procedure (measurement of an absolute index standard) with each prism. Accuracy decreases with film thickness, with measured index systematically low by .001 at 5 microns, and by .003-.004 at 3 microns.

Measurement time: Typical measurement time is 0-20 seconds.

Index measuring range: Minimum and maximum index measurable varies with prism type selected: 1.55-2.45 (200-P-2 prism), 1.25-2.10 (200-P-3), 1.20-2.00 (200-P-4), and 1.00-1.80 (200-P-1). Special 200-P-2 prisms are available to extend measuring range up to approximately 2.60.

Other wavelengths: For index measurement at other wavelengths, the Model 2010/M can be operated with a variety of lasers in place of, or in addition to, the standard 633 nm HeNe. Please consult the factory for details.

Press on chart for larger view

Measurement of ordinary and extraordinary indices of bulk lithium niobate

To top of page

Other Application Notes:

MONITORING OF PHOSPHORUS AND OTHER DOPANT CONCENTRATIONS IN SILICON DIOXIDE
HIGH ACCURACY MEASUREMENT OF RESISTS, POLYIMIDES, AND POLYMERS
DUAL FILM MEASUREMENTS
OPTICAL WAVEGUIDE CHARACTERIZATION

BULK MATERIAL OR THICK FILM INDEX/BIREFRINGENCE MEASUREMENT
MEASURING INDEX ANISOTROPY/BIREFRINGENCE OF FREE-STANDING POLYMER FILMS

 

 

 Home
 Basic Description
 Features & Specifications
 Comparison to Ellipsometry
Journal References
 Applications Overview
 Application Notes
Ordering/Options Guide
Metricon Sales Reps

 

Copyright © 2006 Metricon Corp. All rights reserved.
Last revised: 11/8/06